TECHNICAL FEATURES

Throughput

  • 60 tests/h for TP.
  • 50 tests/h for TP y APTT.

Parameters

  • PT, APTT, TT, FIB (Fibrinogen), ATIII, D-dimer, etc.

Measurement methods 

  • Coagulometric: light scattering method, 660 nm.
  • Chromogenic: colorimetric method, 405 nm.
  • Immunologic: turbidimetric method, 575 nm.

Memory

  • 100.000 tests results and 10.000 reaction curves.

Quality Control

  • 12 Quality Control files. 10 tests per file. 30 curves. 12 months of storage.
  • Dos programas de C.Cal: Levey Jennings (con Reglas de Westgard) y QC Limite.

Calibration

  • Up to 6 points. 10 items.

Type

  • Automatic, random access.

STAT 

  • Yes

Sample auto-dilution and/or re-test

  • Yes

Barcode reader

  • Yes (optional)

Samples tray

  • 27 positions 
  • User-defined STAT.
  • Incubation time: 37±0.5°C

Reagents tray

  • 23 positions
  • Cooling: ≤ 16 °C

Reaction tray

  • Cuvettes on-board: 72
  • Minimum reaction volume: 150μL
  • Reaction temperature: 37 ± 0.5 ° C

Probe

  • Probe preheating.
  • Internal and external automatic washing.
  • Anti-collision system and liquid level detection.
Delivery of results
    • Thermal printer. Optional external printer.
    • Bidirectional connection.
Installation conditions
  • Power supply: AC 110 / 240V, 50-60 ± 1Hz
  • Temperature: 10 ° C – 30 ° C, humidity 85% ≤
  • Water consumption: 0.5 L / h
  • Dimensions: Length / width / height (cm): 79x72x75
  • Weight: 53 kg